LS25
The LS25 laser analyzer selectively measures the concentration of up to two sample components.The laser operates according to the principle of single-line spectroscopy. For measurement purposes a single absorption line is selected from the gas to be measured in the near infrared spectral range, at which no cross-sensitivity from other gases occurs. The absorption line is scanned and the receiver located opposite detects the absorption caused by the sample gas and calculates the gas concentration from this.
LS25激光分析仪可选择测量zui多两种样品组分浓度。激光以单线光谱学原理为基础。测量时,在近红外光谱范围内测量的气体中选择一条单吸收线,此时不会发生来自其它气体的交叉林敏度。对该吸收线进行扫猫,安装在对面的接收器会检测到样气产生的吸收作用,并据此计算气体浓度。
The laser analyzer module consists of a transmission and receiver unit which can be used on ducts with a diameter of 0.5...6 m. Depending on the application, process pressures of 10 bar and temperatures up to1500 °C can be achieved.
浓度该激光分析仪模块配备有一个收发装置,能在管径为0.5...6 m的管道上使用。在实际应用中,过程压力zui高可达10 bar,温度zui高可达1500 °C。
Multianalyzer systems allow the combination of up to four analyzer modules from the product range.
多组分分析系统zui多可以组成同一种产品系列中的四种分析模块。
Typical applications(典型应用邻域) Emission monitoring(排放监测)
Combustion control(燃烧控制)
Process monitoring and analysis in the chemical industry, steel and iron manufacture, power industry and glass manufacture.(化工,钢铁制造,电力和玻璃制造业的过程监测分析)
Control of DeNOx installations(除氮氧化物装置的控制)
测量组分--zui小量程(范例)
O2 0...1 Vol.-%, 10 bar, 1,500 °C
CO 0...3,000 mg/m3, 3 bar, 1,500 °C
CO2 0...3,000 mg/m3
HCl 0...7 mg/m3
H2S 0...300 mg/m3
NH3 0...10 mg/m3
H2O 0...3 mg/m3 low levels
H2O 0...1 Vol.-% high levels
标定
Calibration takes place either via a built-in flow cell or a separate calibrating cell.
标定即可通过内置流动分析池也可通过独立的标定池进行。